LAC Test Results / Actions
Optical Calibration
- Radiometric
- Initial Results Consistent with Models
- Offset Error Limits Dynamic range to 11 bits. Difficult to Differentiate High Albedo from Specular Reflection. 60 Hz Operation Provides Viable Fix.
- Unexpected Temperature Dependence of Quantum Efficiency
- Spectral
- Filters Have Excellent Spectral Characteristics
- Analysis Continues
- Alignment: Within 2 arc-min of ALI Track-- achieved 1
- Flat Field: Solar Calibrators Consistent with Diffuse Illumination